Infrared Temperature range
-40 °C to 800 °C (-40 °F to 1472 °F)
Infrared accuracy
< 0 °C (32 °F): ± (1.0 °C (± 2.0 °F) + 0.1°/1 °C or °F);
> 0 °C (32 °F): ± 1 % or ± 1.0 °C (± 2.0 °F), whichever is greater
Display resolution
0.1 °C / 0.1 °F
Infrared spectral response
8 µm to 14 µm
Infrared response Time
< 500 msec
Thermocouple Type-K input Temperature range
-270 °C to 1372 °C (-454 °F to 2501 °F)
Thermocouple Type-K input accuracy
-270°C to -40°C:
± (1 °C + 0.2 °/1 °C) (-454 °F to -40 °F: ± (2 °F + 0.2 °/1 °F))
-40°C to 1372°C:
± 1 % or 1 °C (-40 °F to 2501 °F: ± 1 % or 2 °F), whichever is greater
D:S (distance to measurement spot Size)
50:1
Laser sighting
Single-point laser < 1 mw Output Class 2 (II) operation, 630 nm to 670 nm
Minimumspot Size
19 mm (0.75 in)
Emissivity adjustment
By built-in table of common Materials or digitAlly adjustable from 0.10 to 1.00 by 0.01
Data storage with Date/Time stamp
99 points
PC Interface and cable
USB 2.0 with FlukeView® Forms software
Hi/Low alarms
Audible and two-Colour visual
Min/Max/Avg/Dif
Yes
Display
Dot matrix 98 x 96 pixels with function menus
Backlight
Two levels, Normal and extra bright For darker environments
Trigger lock
Yes
Switchable Celsius and Fahrenheit
Yes
Power
2 AA/LR6 Batteries and USB when used with a PC
Battery life
If used continuously:
laser and backlight on,12 hours; laser and backlight off, 100 hours
Operating Temperature
0 °C to 50 °C (32 °F to 122 °F)
Storage Temperature
-20 °C to 60 °C (-40 °F to 140 °F)
Bead thermocouple Type-K range
-40 °C to 260 °C (-40 °F to 500 °F)
Bead thermocouple Type-K accuracy
± 1.1 °C (2.0 °F) from 0 °C to 260 °C (32 °F to 500 °F), typicAlly within 1.1 °C (2.0 °F) from -40 °C to 0 °C (-40 °F to 32 °F)